WARNING
= possible injury to patient or operator
CAUTION
= possible damage to equipment
Note
= Provides additional information to clarify a point in the text.
Important
= Similar to a note but of greater emphasis
|O|O (INT)
= intermittent; cycles ON and OFF
| (CONT)
= continuous, ON
O (OFF)
= OFF
= Consult Instructions for Use
= Serial Number
= Manufacturer
Abbreviations
in
Inch
CVR
Continuous Vacuum Regulator
kPa
Kilo pascals (kPa x 7.50 = 1 mmHg)
mmHg Millimeters of mercury
(mmHg x 0.133 = kPa)
mm
Millimeters
mL
Milliliters
°C
Degrees Celsius
°F
Degrees Fahrenheit
WARNINGS
The Pre-Use Checkout Procedure must be performed before using the equipment on each patient. If the
regulator fails any part of the Pre-Use Checkout Procedure, it must be removed from service and repaired
by qualified service personnel.
Do not use this device in the presence of flammable anesthetics. Static charges may not dissipate and a
possible explosion hazard exists in the presence of these agents.
This device is to be used only by persons who have been adequately instructed in its use.
To reduce service personnel exposure to hazardous contamination, clean and disinfect all suction
equipment before disassembly.
To reduce transportation personnel and/or service personnel exposure to hazardous contamination, DO NOT
ship any suction equipment that has been contaminated.
Ohio Medical will assume no responsibility for incidents which may occur if the product is not used in
accordance with product labeling.
Connection to positive pressure sources such as oxygen and medical air, even momentarily, could injure
the patient or operator.
After patient use, regulators may be contaminated. Handle in accordance with your hospital's infection
control policy.
To help prevent aspirate from entering the device, wall outlet and pipeline equipment, a safety trap
should be attached prior to its use. Aspirate in the device, wall outlet and pipeline equipment may impair
its operations. The use of the safety trap and suction filter will help prevent this and extend the life of
suction equipment.

Definitions

N-m
ft-lb
in-lb
DISS
NPTF National Pipe Thread Female (USA)
NEO
PED
PTS
PTFE Polytetrafluoroethylene
8700-0007-000 (Rev. 14) 03/2020
Newton-Meter (N-m x 0.737 = ft-lb)
Foot-Pound Force (ft-lb x 1.356 = N-m)
Inch-Pound Force (ft-lb x 12 = in-lb)
Diameter Index Safety System
Neonatal
Pediatric
Push-To-Set™