5.
Remove the pump head by pulling gently and sliding the pump head off of the motor assembly
6.
Slide the new pump head onto the motor assembly ensuring that the slot in the pump head matches
the notch in the motor shaft. DO NOT force the pump head onto the motor assembly; it should
easily slide into place.
7.
Rotate the pump head and slide the alignment pins into motor assembly
8.
Replace the two (2) screws to reattach the pump head.
9.
Attach the new pump tubes by pressing the tubes onto the connectors on the end of the pump head.
10.
Insert the pump tube leading to the bottom of the pump head into the appropriate reagent bottle
11.
Attach the pump tube from the top of the pump head to the needle on the colorimeter cap by holding
the needle and gently pressing the tube onto the connector.
5.6

Troubleshooting

The following procedures instruct how to locate, evaluate, and fix a problem when HydroGuard issues an
alarm or indicates suspect chemical levels. Table 8 outlines some controller problems that may generate a
false alarm, potential causes and solutions. Table 9 outlines some possible problems associated that may
occur but will not generate an alarm and their potential causes and solutions.
Alarm
#
Not enough water reach
No Flow
A1
Low Flow
A2
Check
Colorimeter
A3
Connect
Stops colorimeter testing
No Reagents
A4
Stuck Piston
A5
Description
the controller. Stop all
chemical dosing (all
relays are open)
Water flow rate in main
circulation pipe is lower
than the flow limit. Stop
all chemical dosing (all
relays are open)
No communication with
the colorimeter
and chlorine dosing.
Displays no CL _ _ . _
value in control panel.
Piston is not moving
properly within the
colorimeter.
Symptom
No reading
board. Check that the switch is
No dosing
No Cl measurements
No reading
Wait for the 2nd measurement
No reading
43
Solution / Suggestion
Increase flow and check flow
switch wire connection on
not stack.
Check the circulation pipes,
not controller problem.
Check wiring Between
colorimeter sensor and
colorimeter board. Change
colorimeter sensor case of
continuing problem.
Check reagent level change
bottle if needed. Make sure
the sensor is not stack.
and check visually. Clean the
colorimeter assembly. Make
sure the colorimeter
connection well connected to
the "mixer" on control panel.
Authorized technician, may
open the cover and check
the coils connection. Try to
push the piston form the top
down