en - Areas of application
Fitting the A 813
32
The A 813 adapter is intended for use on injector manifolds with
screw connections.
Remove a nozzle from the injector manifold of your upper basket or
mobile unit.
Unscrew the top part of the A 813 adapter.
Screw the bottom part of the adapter in at this position.
Screw the top part into the bottom part.
Then screw the nozzles and A 815 silicone hoses into the rinse
ports. Alternatively, you can also connect different rinse fittings
from the Miele range of accessories to the adapter. Ports that are
not needed must be sealed with E 362 blind stoppers.
The wash water used for reprocessing narrow lumen instruments
such as transmission instruments needs to be filtered.
The injector manifold must be fitted with an A 800 tubular filter.
The tubular filter is supplied with its own operating instructions.
Follow the instructions on how to perform regular cleaning on the
tubular filter.