Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

Alligator Pocket/Belt Clip
LCD Display
Audio Alarm Opening
(Buzzer)
110XL-Series Label

SPECIFICATIONS

Size
.2 x 2.0 x .2 in. (8. x . x . cm)
Weight
. oz. ( g)
Ingress Protection Rating
IP
Sensor
Electrochemical (factory replaceable)
Operating Humidity
 – 90% (non-condensing)
Readout
Direct read LCD
Range
0-99 ppm (CO)
0-200 ppm (H
Alarm Indicators
Visual: Flashing LCD display and red LED
Audio: 8 dBA @  foot (0. cm) buzzer
Tactile: Internal vibrator
Alarm Readings
Low, High, Low Battery, Over Range , TWA, and STEL alarms
Alarm Settings
See alarm settings in the table below
Reading Resolution
 ppm (readings start at  ppm)
Test Function
Self-test on circuitry, battery and alarms. Automated sensor test reminder.
Sensor Replacement
Factory replaceable
Intrinsic Safety
CSA (US) - Class I, Div I, Group A, B, C, D, TA
Visual Alarm LED
110XL-CO
(Carbon Monoxide)
110XL-H
S
(Hydrogen Sulfide)
2
On/Off/Menu Button
Sensor Inlet
S)
2
6
SPECIFICATIONS (cont.)
Power
Battery Life
Radio Frequency Protection
Keypad
Operating Temperature
Sensor Accuracy (after
calibration)
Fig. 
Non-contaminated
environment
Certified calibration gas*
Warranty
* A gas analysis certificate of the calibration gas is available if requested.
Gas
Sensor
Read by
Display
Reading
Symbol
Sensor
Range
Resolution
Carbon
0 - 99
CO
Monoxide
ppm
 ppm
Hydrogen
0 – 200
H2S
Sulfide
ppm
 ppm
Using a certified gas with a concentration other than the one listed for this instrument and sensor when
conducting a calibration or calibration verification test (Bump test) will produce inaccurate readings. This
means that higher levels of the gas being monitored may be present and could result in overexposure and
cause sickness or death. For proper use, see supervisor or User Instructions, or call M in U.S.A.,
-800-2-60. In Canada, call Technical Service at -800-267-.
Permanently mounted, non-rechargeable .6- volt Lithium battery (factory
replaceable)
2 years from activation (maximum 6. alarm hours –  minutes/day)
<0% deviation of alarm level when subjected to 0MHz,  Watt Radio
@ 2 ft (6 cm)
One button operation
See table below
Carbon Monoxide (CO) – Less than ± 2% of the displayed reading or
 ppm, whichever is greater
Hydrogen Sulfide (H
S) - Less than ±2% of the displayed reading or
2
2 ppm, whichever is greater
An environment containing less than 0. ppm of hydrocarbons, 0. ppm of
CO, 0.2 ppm of H
S, 0.2 ppm of Ammonia, 0. ppm of Hydrogen and 0.2
2
ppm of other gasses not normally found in the air.
Refers to a calibration gas sold by M or produced by an ISO 900:2000
certified calibration gas manufacturer. The certified accuracy of the
calibration gas components must be listed on the calibration gas container
and traceable to National Institute of Standards and Technology (NIST) or a
national measurement institute standard reference material.
2 years for sensor and electronics (see WARRANTY section)
Required
Factory
Factory
Calibration
Factory
Set
Set
Gas
Set Low
High
TWA
Concentration
Alarm
Alarm
Alarm
200
 ppm
 ppm
ppm
 ppm
0 ppm
0 ppm  ppm 0 ppm  ppm
WARNING
7
Factory
Set
Operating
STEL
Temperature
Alarm
Range
200
- to 22 ˚F
ppm
(-20 to 0 ˚C)
- to 22 ˚F
(-20 to 0 ˚C)

Publicidad

Tabla de contenido
loading

Tabla de contenido