Setup and initial use
Remote control
(Laborota
4001 efficient,
4011 digital,
4003 control)
104
Adjust the depth to which the temperature sensor T auto extends into the
condenser so that the lower edge of the sensor is positioned at 2/3 of the
condenser height.
1. Remove the screw cap at the vacuum connection.
2. Insert the temperature sensor T(auto) into the vacuum connection so that
the white PTFE side (3) of the seal faces the glass thread.
3. Slide the temperature sensor T(auto) to adjust its depth in the condenser.
4. Lock the temperature sensor T(auto) in place with the screw connector
cap (1).
5. Connect the bayonet plug (2) to the socket on the rear of the instrument
(see chapter 4.4.4, page 84).
Fig. 4-28:
Remote control
Connect the bayonet plug of the remote control to the socket on the rear of
the instrument (see chapter 4.4.4, page 84).
Raising and lowering the lift
1. Press the arrow buttons (3) on the remote control.
Starting vacuum/rotation (Laborota 4003 only):
1. Press the Vac Start Stop (2) or Rot Start Stop (1) buttons on the remote
control.
Laborota 4000/4001 efficient, 4010/4011 digital, 4002/4003 control