Input Commands Format; Relay Interface - Adam Equipment ABK Serie Manual De Usuario

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

EN
The RS-232 serial interface is a plug as figure 2 shows:
2
3
The scale can be set to print text in English, French, German or Spanish.
See the RS-232 parameters section of the full user manual for details.
The data format and examples of printouts are shown in the full version of the
user manual.
9.1

INPUT COMMANDS FORMAT

The scale can be controlled with the following commands. Press the [Enter]
key of the PC after each command.
T<cr><lf>
Tares the scale to display the net weight. This is the same as
pressing [Z/T].
Z<cr><lf>
Sets the zero point for all subsequent weighing. The display
shows zero.
P<cr><lf>
Prints the results to a PC or printer using the RS-232
interface. It also adds the value to the accumulation memory
if the accumulation function is not set to automatic.
10.0

RELAY INTERFACE

The indicator is supplied with drivers to control external relays. The drivers
could be used to control a number of different relays depending upon the
users needs. The relay drivers are isolated outputs requiring the use of an
external power supply and the relay option kit.
See of the full version of the user manual for complete details.
© Adam Equipment Company 2015
1
4
As viewed from the back of the indicator
1: Pin GND,
Signal Ground
2: Pin RXD,
Received Data
3: Pin TXD,
Transmitted Data
14
3116611621RevA1-Feb15

Publicidad

Tabla de contenido
loading

Tabla de contenido