52 23 / 52 24
Storage
The connector should be protected from hu-
midity by the protective cap
Place the electrode in the storage protector
with the adequate electrolyte.
52 23: Crisolyt G (KCl + glycerine)
52 24: LiCl Ethanol.
Close the refilling port with the stopper
Life of an electrode
The average life expectancy of a pH elec-
trode is approximately one year depending
on the working conditions.
An electrode used for more than six months
that does not response appropriately after re-
generation, must be replaced by a new one.
To avoid accidental breakage of the glass
body electrodes, the use of the working
protector, code 91 62, is recommended.
If an electrode has too short lifespan,
less than three months, ask us about it.
Probably there is a more suitable electrode
for the application.
Specifications
Measuring range
Operating temp.
Reference
element
Diaphragms
Electrolyte
Body material
Min. immersion depth
Applications
• 52 23: Viscous samples with a high pro-
tein content.
• 52 24: This electrode is basically used to
perform titrations in non-aqueous media.
Limitations
Temperatures > 60 ºC.
Warranty
CRISON guarantees this electrode against
manufacturing defects.
Validity: 6 months from shipment date.
Limitations:
• Accidental damage.
• Inadequate applications.
• Non-fulfi lment of the recommendations.
• Normal wear and tear of the electrode.
QUALITY CERTIFICATE
The attached electrode 52 23 / 52 24 has
passed the quality test and follows the next
specifications:
• Asymmetry potential < ± 15 mV.
• Sensitivity, pH 4...7 (at 25 ˚C) > 98%.
• Response time, pH 4...7 < 20 s.
52 23
0 ... 14 pH
0 ... 60 °C
encapsulated
Ag/AgCl crystals
sleeve
CRISOLYT G LiCl Ethanol
20 mm
GB
.
52 24
glass
.
11