PROBLEM
The appliance switches itself
off or does not come on.
When using the appliance the
battery symbol flashes
The appliance does not come
on and the battery symbol
flashes
The appliance switches itself
off after around 18 minutes.
The timer symbol comes on
and stays on
The appliance switches itself
off and the timer symbol flash
The appliance remains
switched off after pressing
ON/OFF
Nebulization very poor or
absent
Noisy appliance
Sudden nebulization
interruption
DISPLAY
CAUSE
INDICATION
No water or water below the reference
level
The battery is being charged
If you are using the adapter it indicates
a temporary lowering of the power
supply voltage
Flat battery
Timer intervention
The appliance is overheated
Adapter not connected correctly
Green LED
switched off
Battery not connected correctly
Appliance/battery / adapter damaged
Too much medicine (more than 8 ml.) in
the medicine cup
Small amount of medicine
Medicine not inserted
Medicine not suitable for use with
electronic, ultrasound, mesh appliances
Small amount of water in the reservoir
Fan blocked by encrustation
Wet filter / dirty filter
Noisy fan
Air bubble between the cover and
the tube
Wet filter
Medicine has escaped from the
medicine cup due to brusque
movements
SOLUTIONS
Fill the reservoir suitably
Switch of the appliance and charge the battery
You can continue with the therapy, unless the
nebulization becomes insufficient
Charge the battery or power the appliance with
the adapter
Let the appliance cool for at least 30 minutes
Let the nebulizer cool for at least 30 minutes
Check that the adapter is well connected both to
the nebulizer and to the electrical socket
Make correct connection. See the use manual
Contact the stockist or an assistance centre
As indicated in the "USE INSTRUCTIONS"
subdivide the dose
As indicated in the "use instructions" do not insert
less than 4 ml.
Check that medicine is inserted in the medicine
cup
Check the instructions enclosed to the medicine
and consult a doctor
Check that the water arrives at the level indicated
See cleaning instructions
Rinse the filter with water and dry it with a cloth /if
necessary replace the filter
Check that the fan is not encrusted with medicine.
Check that the filter does not touch the fan.
Blow gently into the tube
Rinse the filter with water and dry it with a cloth
Replace the medicine