Position of electrodes: form a square above the painful area as shown in illustration
1.
Intensity adjusted for good solicitation of the part stimulated, the sensation must be
similar to that of a massage.
P05• Medium frequency TENS
Medium frequency TENS with an effect similar to TENS conventional, with the
stimulation of muscular fibers of different diameters. To be used when TENS
conventional does not produce appreciate results.
Treatment duration 30 minutes, frequency 35Hz, impulse width 200μs.
Position of electrodes: form a square above the painful area as shown in
illustration 1.
Adjust the intensity until a tingling feeling is produced in the area treated; avoid
contracting the surrounding muscles.
P06• TENS Burst
This stimulation produces a training TENS effect by using the conventional TENS
frequencies. Commonly used for pain therapy. The action is similar to that of the
endorphinic TENS program, and the treatment time is 30 minutes in a single phase, 4
Hz.
Position of electrodes: form a square above the painful area as shown in illustration 1.
Intensity adjusted for good massage of the part stimulated.
PC1• TENS programmable
Use this program if you need to set you own program, even with the parameters
suggest by your physician.
The treatment time can be setted from 10 to 90 minutes, with adjustable frequency from
2 to 120 Hertz and impulse width from 50 to 400 microseconds. Follow these steps to
set your parameters:
1. Switch on SMART TENS pressing ON/OFF button
2. Select TENS programs using TENS/EMS button (the display shows TENS icon);
3. Select PC1 program and confirm it pressing P button;
4. Keep pressed P button for 3 seconds: the timer icon on the lower right side will start
flashing;
5. Adjust therapy time using
6. Adjust frequency (flashing icon Hz on the lower left side) using
buttons and confirm pressing P button;
7. Adjust impulse width (flashing icon μs on the lower left side) using
buttons and confirm pressing P button;
8. SMART TENS is now ready for the treatment with desired parameters. Press
and/or 2 button (the icon of selected channel will start flash) to chose the channel
and increase intensity pressing
I.A.C.E.R. S.r.l.
(Medical program)
(Medical program)
(Medical program)
and
buttons and confirm pressing P button;
button;
;
37
and
and
1
MNPG45-02