6.6 Timer Measurement
EN
Set measurement time interval 00 hr:00 min in
Menu\Settings\Timer. Press and hold measuring key for 3
DE
seconds to start timer measurement, Timer icon will be shown at
the top left of LCD display to indicate current measurement mode.
FR
To stop timer measurement, press and hold measuring key for 3
ES
seconds.
6.7 Data Transfer
IT
The data is stored in the internal memory D2 and can be
TK
downloaded using the PC based Stonex PC Tools
RU
Data Output
Data via interface and data export in GSI-format.
• GSI-8
8-data characters output format (83..00+12345678).These
data words can hold alpha as well as numeric data.
Communication
Communications parameters of the RS232 interface for data
transfer from instrument to computer / external device.
Stonex standard settings are:
• 19200 baud rate,
• 8 data bits (8 data bits are automatically set when parity "None"
is set.)
• No parity (No parity [when data bits = 8])
• CR/LF (New line and line feed)
• 1 stop bit
Check & Adjust
7. Check & Adjust
7.1 Electronic Collimation Adjustment
A ~15m
To activate "Adjustment" program, go to Menu\Adjustment.
Step 1: Aim at Staff A and press MEAS key. Measurement display,
press ENTER key to accept.
Step 2: Aim at Staff B and press MEAS key. Measurement display,
press ENTER key to accept.
Now shift the D2 towards staff A and set it up at about 3 m to
staff A.
A ~3m
Step 3: Aim at Staff B and press MEAS key. Measurement display,
press ENTER key to accept.
I
B
~15m
~27m
B
*
|
15
D2