EMD Technical Data Battery-operated Blood Pressure Monitor
Medical Electrical Equipment needs special precautions regarding EMD and needs to be installed and put into service
according to the EMD information provided in the following.
Portable and mobile RF communication equipment (e.g. cell phones) can affect Medical Electrical Equipment.
The use of accessories and cables other than those specified may result in increased emissions or decreased immunity
of the unit.
Table 1 -EMISSION Limits-
Phenomenon
Conducted and radiated RF EMISSION
Table 2 -IMMUNITY TEST LEVELS:Enclosure Port-
Phenomenon
Electrostatic discharge
IEC 61000-4-2
Radiated RF EM fields
IEC 61000-4-3
Proximity fields from RF wireless communications equipment
IEC 61000-4-3
Rated power frequency magnetic fields
IEC 61000-4-8
Table 3 -Test specifications for ENCLOSURE PORT IMMUNITY to RF wireless communications
equipment-
Test
Band
frequency
(MHz)
(MHz)
385
380 - 390
450
430 - 470
710
745
704 - 787
780
810
870
800 - 960
930
1720
1845
1700 - 1990
1970
2450
2400 - 2570
5240
5500
5100 - 5800
5785
CISPR 11
Service
TETRA 400
GMRS 460
FRS 460
LTE Band 13、17
GSM 800/900
TETRA 800
iDEN 820
CDMA 850
LTE Band 5
GSM 1800
CDMA 1900
GSM 1900
DECT
LTE Band 1、3、4、25
UMTS
Blutooth
WLAN 802.11 b/g/n
RFID 2450
LTE Band 7
WLAN 802.11 a/n
English 19
Compliance
Group 1、Class B
IMMUNITY TEST LEVELS
±8 kV contact
±2 kV, ±4 kV, ±8 kV, ±15 kV air
10 V/m
80 MHz - 2.7 GHz
80 % AM at 1 kHz
See table 3
30 A/m
50 Hz or 60 Hz
Maximum
Modulation
power
(W)
Pulse modulation
18 Hz
FM
±5 kHz deviation
1 kHz sine
Pulse modulation
217 Hz
Pulse modulation
18 Hz
Pulse modulation
217 Hz
Pulse modulation
217 Hz
Pulse modulation
217 Hz
IMMUNITY
Distance
TEST
(m)
LEVEL
(V/m)
1.8
0.3
27
2
0.3
28
0.2
0.3
9
2
0.3
28
2
0.3
28
2
0.3
28
0.2
0.3
9