This section of the manual is a general introduction to the controls and operation of the instrument
and is intended to be read before using the power supply for the first time.
In this manual front panel keys, connections and display indicators are shown in capitals,
e.g. STORE, ESCAPE, OUTPUT, JOG. Messages shown on the 7-segment display are printed
in a different type-font, e.g.
characters as they are shown on the 7-segment display.
Switching On, Output On/Off
The power switch is located at the bottom left of the front panel.
At power-up the default behaviour is for the instrument's settings to be restored to those
automatically saved when it was switched off, but with all outputs always off. However, the user
can change the default setting such that selected outputs are restored at power-up to their status
at power-down, see the Extra Functions section.
The DC outputs are switched electronically with their respective ON/OFF keys; the key illuminates
when the output is on. In addition, all outputs can be switched on and off together using the ALL
ON and ALL OFF keys.
Synchronous Output On/Off Switching (T models only)
Pressing ALL OFF at any time will synchronously turn off any outputs that are on; under the same
load conditions outputs will typically turn off within 1ms of each other. With all outputs off the ALL
OFF key is illuminated green. Pressing the ALL ON key when all the outputs are off will turn all
the outputs on synchronously; outputs with identical settings and load conditions will typically turn
on within 1ms of each other. However, if one output is already on, pressing ALL ON will turn the
remaining outputs on but the turn-on delay between the outputs will be up to 80ms, even with the
same output setting and load conditions.
Keypad
Only the principles of operation are outlined here; the setting of individual parameters is given in
detail in later sections.
The paramount consideration in designing the user interface has been to make changing settings
as 'safe' as possible (i.e. with minimal risk of accidentally applying excessive voltages to a target
system) whilst achieving ease of use. This has been achieved by requiring the user to confirm
(OK) new numeric settings, with the option to ESCAPE at any point or even to simply pause until
the operation times-out and the instrument returns to its original settings.
In addition a buzzer, illuminated keys, LED indicators and display messages prompt, guide or
warn the user such that entry or control errors are minimised. Where some of these features (e.g.
beeps or flashing indicators) are considered unnecessary by regular users, the option exists to
disable them, see the Extra Functions section.
On triple output (T) models the ability to change settings from the keypad or by using the Jog
controls is assigned to Output 1, Output 2 or both by using the 1, 2 or LINK CONTROL keys
respectively. The key (1 or 2) associated with the selected output illuminates to show which
output is under control. In LINK mode (both keys lit) both outputs are controlled at the same time,
including some of the shifted operations (RANGE, STORE, RECALL and V x A). The further
descriptions that follow apply to either or both main outputs as appropriate to the setting indicated
by the illumination of the CONTROL keys.
Under normal conditions the numeric keypad is disabled; pressing any key will cause the buzzer
to make a double beep, indicating an illegal operation. To set a voltage or current with the keypad
press the V or I NUMERIC SET key; the appropriate display shows 0·000V or 0·000A with the
digit to the left of the decimal point flashing. Digits are entered in response to the flashing prompt,
together with the decimal point at the appropriate time, and the entry is confirmed with the OK
key. If OK is not pressed within 10 seconds of the last numeric key the entry is cancelled and the
display returns to its original setting. If ESCAPE is pressed anywhere in the entry procedure,
entry is cancelled and the display returns to its original setting.
StorE, GPIb, triP

Initial Operation

in upper or lower case to represent the