- When the measuring current >1A, press
frequency measurement mode. But the frequency readings obtained from this range is only for
reference.
- Change to AC:
Couple AC and True RMS response method. Input sine wave to adjust. Non sine wave must
follow the below data to adjust:
Peak factor: 1.4~2.0, add 1.0% on the stated accuracy.
Peak factor: 2.0~2.5, add 2.5% on the stated accuracy
Peak factor: 2.5~3.0, add 4.0% on the stated accuracy.
A. General Specifications:
- Display: 3 3/4 digits LCD display, Maximum display 3999.
- Polarity: Auto
- Overloading: Display OL or –OL.
- Battery Deficiency: Display
- Sampling: 3 times per second.
- Measurement Deviation: The conductor being meaured is not placed in the center of the jaw
during AC/DC current measurement, it will cause extra ±1% deviation based on the stated
accuracy.
- Drop Test: 1 meter drop test passed.
- Max. Jaw Size: 55mm diameter.
- Projected Max. Current conductor size: 45mm diameter.
- Electro-Magnetic: When carrying out measurement near the electromagnetic, it may cause
unstable or wrong reading.
- Power: 1 x 9V battery (6LF22 1604A)
- Dimensions: 285.3mm x 105mm x 44.5mm
- Weight: Approximate 533g (battery included)
B. Environmental Requirements
- The Meter is suitable for indoor use.
- Altitude: Operating: 2000m
Storage: 10000m
- Safety/ Compliances: IEC 61010 CATII 600V, CATIII 300V over voltage and
double insulation standard, pollution degree 2.
- Temperature and humidity:
Operating:
0ºC~30ºC(≤85%R.H)
30ºC~40ºC (≤75%R.H)
40ºC~50ºC(≤45%R.H)
Storage:
-20ºC~+60ºC (≤85%R.H)