Descargar Imprimir esta página

Laerdal VitalSim Instrucciones De Uso página 10

Ocultar thumbs Ver también para VitalSim:

Publicidad

Idiomas disponibles

Idiomas disponibles

Set Waiting Rhythm
From the main display use the left / right navigation buttons to
highlight the ECG functional area. Scroll down to Waiting rhyt-
hm. Press OK to edit the selection.
Select a Waiting rhythm in the same way
as the Running rhythm.
When a Waiting rhythm is selected, two additional quick
execution buttons appear in the quick execution area:
• Activate Activates the Waiting rhythm, substituting
the Running with the Waiting.
• Parox. Activates the Waiting rhythm, and exchanges
the Waiting and Running rhythm.
Set Defibrillation Enabled / Disabled
When Defib. Enabled is selected the Waiting rhythm will
be automatically activated and replace the Running rhythm
when a shock is delivered to the manikin.
Enable /Disable is selected by toggling OK or
the quick execution button labeled Disable / Enable
Pacing Threshold
If pacing is enabled pacing pulses on the manikin with a current
higher or equal to the set threshold will activate heartbeats.
Pacing threshold can be adjusted directly from the main display
using the Thres. +/- quick execution buttons when the Pacing
field is selected.
Press OK to set the pacing threshold in a sub menu.
Threshold can be adjusted using the up/ down navigation
buttons, the Thres.+/- keys (steps of 10) or entered directly
using the numeric buttons.
10

Publicidad

loading

Productos relacionados para Laerdal VitalSim