10. Technical Alarm Description

The appliance will show "HI" or "Lo" as technical alarm on LCD with no delay if
the determined blood pressure (systolic or diastolic) is outside the rated range
specified in part "Specifications". In this case, you should consult a physician or
check if your operation violated the instructions.
The technical alarm condition (outside the rated range) is preset in the factory
and cannot be adjusted or inactivated. This alarm condition is assigned as low
priority according to IEC 60601-1-8.
The technical alarm is non-latching and need no reset. The signal displayed on
LCD will disappear automatically after about 8 seconds.

11. Troubleshooting (1)

Problem
Possible cause
The cuff position was
not correct or it was not
properly tightened.
Body posture was not
LCD Display
correct during testing.
shows
abnormal
Speaking, arm or body
result
movement, angry, excited
or nervous during testing.
Irregular heartbeat
(arrhythmia)
114
PC-BMG3019_IM
Solution
Apply the cuff correctly and try
again.
Review the section "Body
posture during measurement" of
the instructions and try again.
Re-test when calm and without
speaking or moving during the test.
It is inappropriate for people with
serious arrhythmia to use this elec-
tronic sphygmomanometer.