Technical Data
Internal measuring gages
mm
Measuring range
[mm]
Scale interval
[mm]
Max. permissible errors „G"
[mm]
Repeatability limit „r"
[mm]
Measuring force
[N]
Type of measuring contact
[mm]
Internal measuring gages
inch
Measuring range
[inch]
Scale interval
[inch]
Max. permissible errors „G"
[inch]
Repeatability limit „r"
[inch]
Measuring force
[N]
Type of measuring contact
[inch]
Reference temperature
[°C]
Working temperature
[°C]
Storage temperature
[°C]
Protection class
Technical data sheets for each gage are available at www.kroeplin.com
All gages are available in digital with interfaces and IP67!
Factory setting is checked subject to VDI/VDE/DGQ 2618 Part 13.1
H2G20
H2G30
H2G40
20 – 40
30 – 50
40 – 60
0,01
0,01
0,01
0,03
0,03
0,03
0,01
0,01
0,01
1,1 – 1,6
1,1 – 1,6
1,1 – 1,6
Chisel-R 0,2
Chisel-R 0,2
Chisel-R 0,2
H7G20
H7G30
H7G40
.80 – 1.60
1.20–2.00
1.60–2.40
.0005
.0005
.0005
.0015
.0015
.0015
.0005
.0005
.0005
1,1 – 1,6
1,1 – 1,6
1,1 – 1,6
Chisel-R .01
Chisel-R .01
Chisel-R .01
H2G50
H2G60
H2G70
50 – 70
60 – 80
70 – 90
0,01
0,01
0,01
0,03
0,03
0,03
0,01
0,01
0,01
1,1 – 1,6
1,1 – 1,6
1,1 – 1,6
Chisel-R 0,2
Chisel-R 0,2
Chisel-R 0,2
H7G50
H7G60
H7G70
2.00–2.80
2.40–3.20
2.80–3.60
.0005
.0005
.0005
.0015
.0015
.0015
.0005
.0005
.0005
1,1 – 1,6
1,1 – 1,6
1,1 – 1,6
Chisel-R .01
Chisel-R .01
Chisel-R .01
20
+10 to +30
-10 to +50
IP 65
H2G80
H2M50
H2M90
80 – 100
50 - 100
90 - 140
0,01
0,01
0,01
0,03
0,03
0,03
0,01
0,015
0,015
1,1 – 1,6
1,1 – 1,6
1,1 – 1,6
Ball ø1
Ball ø1
Chisel-R 0,2
H7G80
H7M50
H7M90
3.20–4.00
2.00–4.00
3.60–5.60
.0005
.0005
.0005
.0015
.0015
.0015
.0005
.001
.001
1,1 – 1,6
1,1 – 1,6
1,1 – 1,6
Ball ø.04
Ball ø.04
Chisel-R .01
H2M130
130 - 180
0,01
0,03
0,015
1,1 – 1,6
Ball ø1
H7M130
5.20–7.20
.0005
.0015
.001
1,1 – 1,6
Ball ø.04