Clearing The Memory - Nutricia Flocare Infinity Instrucciones De Uso

Flocare infinity
Ocultar thumbs Ver también para Flocare Infinity:
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

Ensure that the volume of nutrition in the container is greater than is actually needed, i.e. more than the set
DOSE to be administered, so as to avoid excessive air bubbles getting in the feeding line.
d
ON/OFF
ml/h
From the hold mode, with the last flow rate (ml/h) displayed in the LCD:
• Set the flow rate by pressing the "+" or "-" key.
If another parameter is displayed:
• Press first the "ml/h" key followed by "+" or "-" key to adjust the flow rate. The flow rate can be set between
1 and 400 ml/h, with increments of 1ml.
• Press the "DOSE=VOL" key and install the volume to be administered with the "+" and "-" key. The range
goes from 1 - 4000 ml, with steps from 1ml.
• Start the pump by pressing "START/STOP".
During functioning the following parameters can be found in the display:
• By pressing the "ml/h" key the "flow rate" is displayed.
INFO
When the required volume or dose has been delivered, "END OF DOSE" will appear and the pump will beep
(or mute, depending upon the pumps' configuration (see paragraph: set up mode).
• Press the "START/STOP" key to pause the pump.
• Change the program by pressing the required key (ml/h or DOSE=VOL) and adjust using the "+" or "-" keys.
• Restart the pump by pressing the "START/STOP" key again.
All parameters and values can be cleared one after the other in the following way:
• Press "START/STOP" to pause the pump.
• Select the parameter that needs to be cleared.
• Press the "CLR" key to clear the memory. The parameter or value returns to its default value:
ml/h = 0 ml/h
VOL = cont
= No dose set. Pump will feed continuously until feed container is empty or pump
INFO = 0 ml
= Volume delivered is cleared
In order to keep a clear overview of the daily amount of feed delivered, clear the total volume delivered as
each daily feeding schedule is started, as follows:
START/STOP
CLR
ZMNL900616_AW.indd 20
TO SET A DOSE
+
400 ml/h
d
d
DOSE=VOL
-
1 ml/h
• By pressing the "INFO" key the total volume since the memory was cleared will be
visualized.
TO CHANGE THE FEEDING PROGRAM DURING FUNCTIONING

CLEARING THE MEMORY

is stopped
• Switch the pump "ON"
• Immediately after the self-test the pump displays total volume delivered.
• Press the "CLR" key. The total volume delivered returns to "0 ml".
• The pump displays the previous programmed flow rate (ml/h).
• The pump is in the "HOLD" mode ready for programming.
UK-IE
+
4000 ml
d
d
START/STOP
-
1 ml
-20-
10/11/16 16:07

Hide quick links:

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

35679

Tabla de contenido