Idiomas disponibles

  • ESPAÑOL, página 63
28
5. Discharge sample
a) Place the pipette tips against the wall of the vessel. Hold the
pipette at an angle of 30-45° relative to the container wall.
b) Press the pipetting button slowly to the first stop and hold it
down. For serum and liquids of high viscosity or low surface
tension, observe adequate waiting time to improve accuracy.
c) The blow-out stroke empties the tips completely: Press the
pipetting button down to the second stop.
d) While doing this, wipe the pipette tips against the wall of the
container.
e) Remove the pipette tips from the container wall and let the
pipetting button slide back.
6. Ejecting the tips
Hold the manifold over a suitable disposal container and press the
tip ejection key to the stop.
Note:
ISO 8655 prescribes rinsing the pipette tips once with the
sample liquid prior to the actual pipetting process.

Important!

Don't lay the instrument horizontal when the tips are filled.
Liquid may enter and contaminate the instrument.