SpeCiFiCation
Temperature Range
Accuracy
Repeatability
Response Time
Spectral Response
Emissivity
Temperature display
Ambient operating range
Relative Humidity
Weight
Dimensions
Power
Laser
CaLeXSoFt SoFtWare
SySTEm REquiREmENTS
•
CalexSoft software is designed to run under Windows. Suitable versions are Windows
2000, Windows XP and Windows Vista.
•
The recommended minimum computer specification is:
Intel Pentium processor
VGA display with 640x480 resolution
USB 2.0 port
iNSTallaTiON
1. Insert the disk provided
2. Installer should run automatically (if not, select the appropriate drive and double click
on Setup.exe)
3. Follow the on-screen instructions
NOTE Do not connect the PyroPen u to the PC before the software is installed
CONNECTiNg ThE PyROPEN u TO a PC
1. Connect the USB Mini B connector on the USB cable provided to the PyroPen U.
2. Connect the USB A connector on the USB cable provided to an available USB 2.0
port
3. Double click on the CalexSoft desktop icon
NOTE Do not disconnect the uSB cable before closing down the CalexSoft
program
ConFiGUration
•
Setting Temperature Units
Temperature units can be set on the main temperature panel by pressing °C or °F.
•
Setting Target Emissivity
To set an Emissivity value for the PyroPen U, use the main program menu to select:
[Setup] → [Emissivity]
Emissivity can then be set to a value between 0.3 and 1.0, or selected from a pre-
set list of materials by selecting [Data].
-20°C to 500°C
±1% or ±1°C (±2°F) whichever is greater
±0.5% of reading or ±0.5°C (±1°F) whichever is greater
500mSec 95% response
8 to 14µm
adjustable 0.3 to 1.0
0.1°C
0 to 50°C (32 to 120°F)
10-95% RH non-condensing
60g
163 x 27 x 16mm
2 x AAA batteries
Class 2
5