Fault
Cause
No connection via
Cable faulty
cable to radio
Cable not fitted
possible
correctly
Incorrect cable
connected
Pressure switch jams Pressure switch
soiled
Slide switch jams
Slide switch soiled
Slide switch damaged Replace slide switch.
Dräger FPS-COM 7000
5
Remedy
Replace cable.
5.1
Check connection
between cable and
radio. Fit cable again
Work to be carried out
if necessary.
Check connection
Check the functioning of the communication unit
between cable and
communication unit.
5.2
Fit cable again if
necessary.
Use suitable cable.
!
Clean pressure
switch.
Dismantle and clean
slide switch.
i
i

Maintenance

Maintenance intervals

Cleaning

CAUTION
Health hazard!
The undiluted agents are harmful to health if they come into
direct contact with eyes or skin. Wear goggles and protective
gloves when working with these agents.
Potential damage to component parts!
For cleaning and disinfecting only apply the described
procedures and use the specified cleaning agents and
disinfectants. Other agents and procedures, dosages and
exposure times can damage component parts.
NOTICE
The specified cleaning agents and disinfectants are not
approved in all countries. Dräger tests and recommends
different cleaning agents and disinfectants. Information on
which other cleaning agents and disinfectants can also be
used is available from Dräger.