Electromagnetic Compatibility - Air Liquide Soffio Cube Instrucciones De Uso

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 18
COMPONENT
Compressor unit
External casing of the compressor unit
Nebulizer and connector
Mouthpiece
Nosepiece
Masks
Connecting tube
Box
Instructions for use
Bag
Air filters

8 - ELECTROMAGNETIC COMPATIBILITY

This device complies with the requirements of EN 60601-1-2:2015 on the electromagnetic
compatibility of medical devices. Compliance with electromagnetic compatibility standards
does not guarantee total immunity of the product. Some devices (mobile phones, pagers,
etc.) may interrupt operation if used near medical devices.
RECOMMENDED DISTANCES BETWEEN RADIO-COMMUNICATION DEVICES
The SOFFIO CUBE aerosol therapy device is designed to operate in an electromagnetic
environment where radiated RF disturbance is under control. The customer or user of the
SOFFIO CUBE aerosol therapy device can help prevent electromagnetic interference by
leaving a minimum distance between mobile and portable RF communication devices
(transmitters) and the SOFFIO CUBE aerosol therapy device. The table below shows
the recommended distances for the SOFFIO CUBE aerosol therapy device based on the
maximum output of the radio communication devices.
Nominal
maximum
transmitter output
[W]
0,01
0,1
1
10
100
GLISH • ENGLISH • ENGLISH • ENGLISH • ENGLISH • ENGLIS
Separation distance [m] at the transmitter frequency
150 kHz to 80 MHz
d = 1,2 x √P
0,12
0,38
1,2
3,8
12
MATERIAL
Waste from Electric and Electronic
Equipment (WEEE)
Polypropylene
Polypropylene
Polypropylene
Polypropylene
Non toxic PVC
Non toxic PVC
Cardboard
Paper
Nylon
Needled polypropylene
80 MHz to 800 MHz
d = 1,2 x √P
0,12
0,38
1,2
3,8
12
800 MHz to
2,5 GHz
d = 2,3 x √P
0,23
0,73
2,33
7,3
23

Publicidad

Tabla de contenido
loading

Tabla de contenido