3. Specications
• General Specications
Meas. Ranges
Response Time
Over Range
Indication
Polarity
Emissivity
Field of view
Diode Laser
Spectral Response
Power off
Operating Temp
Storage Temp.
Relative Humidity
Power Supply
Weight
Size
• InfraRed Thermometer Specications
Range
-30°C to 0°C (-22°F to 32°F)
0°C to 260°C (32°F to 500°F)
Note:
Accuracy is given at 18°C to 28°C (64°C to 82°C), less than 80%RH.
Emissivity:
0.95 xed value
Field of View:
Make sure that the target is larger than the unit's spot size.
The smaller the target, the closer you should be to it. When
accuracy is critical, make sure the target is at least twice as
large as the spot size.
-30°C to 260°C / -22°F to 500°F
Less than 1 second
LCD will show "---- "
Automatic (no indication for positive polarity). Minus (-)
sign for negative polarity.
0,95 Fixed value
D/S = Approx 8:1 ratio (D = distance, S = spot)
(Has 90% encircled energy at the focal point)
Output: < 1 mW, Wavelength : 630-670 nm, class 2(II)
Laser product
6-14 μm
Automatic shut off after 8 seconds, approx.
0°C to 50°C / 32°F to 122°F
-20°C to 60°C / -4°F to 140°F
10% to 90% RH operating, < 80% RH storage
9V battery, NEDA 1604A or IEC 6LR61 or equivalent
180 g
82 x 41.5 x 160 mm
Resolution
0,1°C/°F
Accuracy
±4°C/±7°F
±2% of reading or ±2ºC/±4°F