Cleaning The Hydrim C51Wd; Filter And Wash Arm Maintenance - SciCan HYDRIM C51wd Manual Del Operador

Ocultar thumbs Ver también para HYDRIM C51wd:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

6. Care and Maintenance

6.4 Cleaning the Hydrim C51wd

To clean outer surfaces and the decal covering the touchscreen, use a soft cloth
moistened with soap and water or a surface disinfectant. Do not use harsh cleaning
chemicals.
If preparing the unit for transport or storage, waste water should be removed from the
sump.
Waste water in the unit may contain biological contaminants. Use a mechanical means to
siphon the contents. Wear disposable rubber gloves. Dispose of absorbent material
according to biological waste disposal regulations.

6.5 Filter and Wash Arm Maintenance

Fine filter – beneath coarse filter
HYDR IM C51wd Operator's Manual
Ring
Filter Maintenance:
Inspect the coarse and fine filters in the
bottom of the chamber daily for debris and
clean if necessary. To clean, remove the filter
(turn the ring to release it), rinse under a
tap and reassemble. Ensure that the filter is
firmly locked into position when replaced.
Coarse filter
Upper wash arm
Wash Arm Maintenance:
If you see that the wash arms are not
turning easily, remove them. To remove the
upper arm, unscrew the collar. To remove
the lower arm, grasp both ends and pull
up. Rinse under a tap and reassemble.
Lower wash arm
Page 22

Publicidad

Tabla de contenido

Solución de problemas

loading

Tabla de contenido