(approximately 40°C) with a gentle, non abrasive dish detergent.
method B: Sanitise accessories C1.1-C1.2-C1.3-C1.4-C2-C3 in the dishwasher with a hot
cycle.
ns,
method C: Sanitise accessories C1.1-C1.2-C1.3-C1.4-C2-C3 by immersing them in a
try
solution of 50% water and 50% white vinegar, then rinse thoroughly under potable hot
for
water (approximately 40 °C).
If you want to also perform the cleaning for DISINFECTION, jump to the DISINFECTION
paragraph.
After having sanitised the accessories, shake them vigorously and place them on a paper
on
towel. Alternatively, dry them with a jet of hot air (for example, a hair dryer).
DISINFECTION
After sanitising the nebuliser cup and the accessories, disinfect them choosing one of the
methods described below.
our
method A: Accessories C1.1-C1.2-C1.3-C1.4-C2-C3 can be disinfected.
on,
The disinfectant must be an electrolytic chloroxidizer (active principle: sodium
mal
hypochlorite) specific for disinfecting, which is available in any pharmacy.
in
Implementation:
- Fill a container big enough to hold all of the parts to disinfect with a solution of potable
water and disinfectant, according to the proportions indicated on the packaging of the
disinfectant.
- Completely immerse each part in the solution, taking care to avoid the formation of air
oft
bubbles on the parts. Leave the parts immersed for the amount of time indicated on the
ad-
packaging of the disinfectant associated with the concentration chosen for the solution.
an
- Remove the disinfected parts and rinse abundantly with warm potable water.
dis-
- Dispose of the solution following the instructions provided by the disinfectant
ater
manufacturer.
lso
method B: Sanitise the accessories C1.1-C1.2-C1.3-C1.4-C2-C3 by boiling them in water
for 10 minutes; use demineralised or distilled water to prevent calcium deposits.
method C: Sanitise the accessories C1.1-C1.2-C1.3-C1.4-C2-C3 with a hot steam steriliser
the
for baby-bottle (not the microwave type). Perform the process faithfully following the
mit-
instructions of the steriliser. To ensure that the disinfection is effective, choose a steriliser
om
with an operating cycle of at least 6 minutes.
After having disinfected the accessories, shake them vigorously and place them on a
paper towel. Alternatively, dry them with a jet of hot air (for example, from a hair dryer).
REPLACING THE FILTER
the
The device has a suction filter (A3) that should be replaced when it is
dirty or changes color. Do not wash or reuse the same filter. Regular
any
replacement of the filter is required to help and ensure correct
performance of the compressor. Check the filter on a regular basis.
Contact your local dealer or authorized service center for spare filters.
zzle
To replace the filter:
bly
Insert a flathead screwdriver between the edge of the filter and the
body. Lift the filter and remove it by turning and pulling it. The filter
has been designed so that it is always fixed in its seat.
Use original FLAEM accessories only.