4. Maintenance
4.1
General Maintenance Procedures
The performance and safety of the instrument will be kept if the following recommendations for care and maintenance are
observed:
The instrument must go through at least one annual overhaul, to ensure that the acoustical, electrical and
•
mechanical properties are correct. This should be made by an authorised workshop in order to guaranty proper
service and repair as Radioear provides the necessary circuit diagrams etc. to these repair shops.
To ensure that the reliability of the instrument is kept, it is recommended that the operator at short intervals, for
•
instance once a day, perform a test on a person with known data. This person could be the operator him/herself.
After each examination of a patient, it should be ensured that there is no contamination on the parts in connection
•
with the patient. General precautions must be observed in order to avoid that disease from one patient is
conducted to others. If ear cushions or ear tips are contaminated, it is strongly recommended to remove them from
the transducer before they are cleaned. By frequent cleaning water should be used, but by severe contamination it
may be necessary to use a disinfectant. The use of organic solvents and aromatic oils must be avoided.
4.2 How to clean Radioear products
Before cleaning always switch off and disconnect from the power supply
•
Use a soft cloth lightly dampened with cleaning solution to clean all exposed surfaces
•
Do not allow liquid to come in contact with the metal parts inside the earphones / headphones
•
Do not autoclave, sterilize or immerse the instrument or accessory in any fluid
•
Do not use hard or pointed objects to clean any part of the instrument or accessory
•
Do not let parts that have been in contact with fluids dry before cleaning
•
Rubber ear-tips or foam ear-tips are single use components
•
Ensure that isopropyl alcohol does not come into contact with any screens on the instruments
•
Ensure that isopropyl alcohol does not come into contact with any silicone tubes or rubber parts
•
Recommended cleaning and disinfection solutions:
Warm water with mild, nonabrasive cleaning solution (soap)
•
Normal hospital bactericides
•
70% isopropyl alcohol only on hard cover surfaces
•
Procedure
Clean the instrument by wiping outer case with a lint free cloth lightly dampened in cleaning solution.
•
Clean cushions and patient hand switch and other parts with a lint free cloth lightly dampened in cleaning solution.
•
Make sure not to get moisture in the speaker portion of the earphones and similar parts
•
4.3 Concerning Repair
Radioear is only considered to be responsible for the validity of the CE marking, effects on safety, reliability and performance
of the equipment if:
1.
assembly operations, extensions, readjustments, modifications or repairs are carried out by authorised persons,
2.
a 1 year service interval is maintained
3.
the electrical installation of the relevant room complies with the appropriate requirements, and
4.
the equipment is used by authorised personnel in accordance with the documentation supplied by Radioear.
AP70 Instructions for Use - English