Troubleshooting - HoMedics BPW-201 Manual De Usario

Monitor automático de presión sanguínea para la muñeca
Ocultar thumbs Ver también para BPW-201:
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

TROUBLESHOOTING

If any abnormality arises during use, please check the following points:
SYMPTOMS
Unit does not turn on
when the START/STOP
button is pushed.
EE measurement error
symbol shown on display
or the blood pressure
value is displayed exces-
sively low
(or high).
E1 error symbol shown
on display
E2 error symbol shown
on display
E3 error symbol shown
on display
EP error symbol shown
on display
Note: If the unit still does not work, contact HoMedics Consumer Relations. Under no circumstance
should you disassemble or attempt to repair the unit by yourself. Contact information for
HoMedics Consumer Relations Department can be found on the warranty page.
POSSIBLE
CAUSES
Batteries have
run down.
Battery polarities have
been positioned incor-
rectly.
Cuff has been placed
incorrectly.
Did you talk or
move during
measurement?
Shaking of the wrist with
the cuff on.
Air circuit
abnormality.
Inflation pressure exceeding
300 mmHg.
Error determining mea-
surement data.
System error.
33
CORRECTION
Replace them
with two new "AAA" alkaline
batteries.
Re-insert the
batteries in the
correct positions.
Wrap the cuff properly so that it
is positioned correctly. Measure
again.
Keep wrist steady during
measurement. Measure again.
Measure again.
Switch the unit off, then mea-
sure again.
Measure again.
Measure again.

Publicidad

Capítulos

Tabla de contenido
loading

Tabla de contenido