TURNING ON THE THERMOMETER
The device must be installed and put into operation
in accordance with the instructions in this manual.
To get help concerning usage or maintenance plea-
se contact your dealer or the manufacturer.
An unexpected operating condition or an incidence
which has worsened the health condition or could
have worsened the health condition should be
reported to the manufacturer immediately.
To turn on the thermometer, push the ON/OFF-button
next to the display; a short signalling tone [beep] in-
dicates «thermometer ON.» A so-called LCD-test
automatically takes place to check the display. Then
a «L» and a flashing «°C» appear on the upper right of
the display, if the ambient temperature is less than
32°C. Now the thermometer is ready for use.
USING THE THERMOMETER
During a measurement, the temperature is displayed
continuously and the «°C» symbol flashes. If the sig-
nalling tone sounds (beep-beep-beep-beep 10 times)
and the «°C» no longer flashes the measurement is
completed (the measured increase in temperature is
now less than 0,1 °C in 8 seconds).
To extend the lifetime of the battery, turn off the ther-
mometer after usage by pressing the ON/OFF-button
briefly. Otherwise the thermometer will automatically
turn off after about 10 minutes.
MEMORY OF MEASURED VALUES
If you hold the ON/OFF-button down for more than 3
seconds when turning the thermometer on, the tem-
perature of the last measurement will be displayed.
At the same time, a small «M» for memory will appear
on the right of the display. After 3 more seconds, this
value will disappear and the thermometer will switch
to the normal measurement mode. Then the stored
value is erased.
CLEANING AND DISINFECTING
For cleaning of the instrument please use isopropyl
alcohol (70%). For disinfecting wipe the instrument
completely with Antifect Liquid (please obey 5 min.
work-in-time).
CHANGE OF BATTERY
As soon as the symbol "
the battery is weary and a new battery has to be in-
serted. Open the cover of the battery compartment
and insert the new battery with "+"symbol facing up-
wards.
" appears on the display,
M M
TECHNICAL SPECIFICATIONS
Maximum thermometer
Measuring range: 32.0°C to 42.9°C
Temp.< 32.0°C:display «L» for low (too low)
Temp.> 42.9°C:display «H» for high (too high)
Accuracy: ± 0.1°C between 34°C and 42°C at an am-
bient temperature of 18°C to 28°C
Operating conditions: environmental
+10°C to +40°C, relative humidity 15 to 95%
Self-test: automatic internal check at a test value of
37.0°C; if there is a deviation of >0.1°C, «ERR» (error)
is displayed.
Liquid crystal display (LCD) with three digits. Smallest
unit of display: 0.1°C
Signalling tone sounds when thermometer is ready
for use and when measurement is completed.
Memory: automatic memory of the last measured va-
lue.
Storage conditions:
25°C to +60°C, relative humidity 15 to 95%
Battery: 1.5/1.55 V
1,000 measurement cycles (depending on frequency
of use).
Type BF applied part
IEC 60417-5333
T
YPE BF APPLIED PART
Norm: EN12470-3, clinical thermometers
Expected operational lifetime of the equipment: 10
years
WARRANTY
This unit is under warranty for 2 years from the date
of purchase.
Please submit the receipt, invoice or delivery note
when claiming for warranty.
For both, warranty and repair, please send the instru-
ment carefully packed and postage paid either to
your authorized dealer or directly to
BOSCH + SOHN GMBH U. CO. KG
Bahnhofstr. 64
72417 Jungingen – Germany
Warranty and repair work must be undertaken by trai-
ned and authorised personnel. Do not modify this
equipment without authorization of the manufacturer.
CARE FOR THE ENVIRONMENT
Used batteries and defective electronic devices do
not belong into the household waste.
They can be deposited at special centres. Please
contact your local authorities for information.
temperature
environmental temperature -