Cleaning And Maintenance; Preventive Maintenance - Arjohuntleigh OBT0 Instrucciones De Uso

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 58

5. Cleaning and Maintenance

5.1
Cleaning
Wearing suitable protective clothing, clean all surfaces with a disposable cloth soaked in a
neutral detergent and hand hot water.
Start by cleaning the upper sections of the Overbed Table and work along all horizontal
surfaces. Work methodically towards the lower sections of the Overbed Table and clean
the castors last. Take extra care to clean areas that may trap dust or dirt.
Rinse with clean water and dry with disposable paper towels.
5.2
Disinfecting
After cleaning the Overbed Table as described above, wipe all surfaces with sodium
dichloroisocyanurate (NaDCC) at a concentration of 1,000 parts per million (0.1%) of
available chlorine.
In the case of pooling body fluids, e.g. blood, the concentration of NaDCC should be
increased to 10,000 parts per million (1%) of available chlorine.
Avoid excessive wetting of the product, particularly the table top
Do not use abrasive compounds or pads
Do not use phenol-based disinfectants
5.3

Preventive Maintenance

The following preventive maintenance procedures should be carried out at intervals of 12
months. All service or repair activities must be carried out by properly qualified and
trained persons approved by ArjoHuntleigh.
Check that the height adjustment mechanism functions smoothly and that the main table
locks securely in position over the full range of Overbed Table height.
Make sure all four castors are securely fixed to the base and that they swivel and rotate
freely. Remove any debris that may have been picked up by the wheels.
Models OBT1 & OBT2 only: Check that the main table locks securely at each tilt position.
Caution
13
416290-INT1 Rev A

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

Obt2

Tabla de contenido