NSK DynaLED M500LG M4 Manual Del Operador página 13

Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

Autoclave Procedure:
1) Clean and lubricate the handpiece.
2) Insert into an autoclave pouch. Seal the pouch.
3) Autoclavable up to max. 135˚C.
ex.) Autoclave for 3 min. at 134°C.
4) Keep the handpiece in the autoclave pouch to keep it clean until you use it.
* EN13060/EN ISO17665-1 recommends autoclaving for 3 minutes (minimum holding time) at 134°C.
NSK recommends Class B sterilization.
CAUTION
• To prevent infectiveness, Autoclave the handpiece every patient.
• Do not autoclave the handpiece, even when it is bagged, together with other instruments, from which chemicals are not sufficiently cleaned.
This is to avoid discoloration and/or damage to the plating of the handpiece.
• In case the sterilizer chamber temperature could exceed 135°C during the dry cycle, skip the dry cycle.
• Do not heat or cool the handpiece quickly. Rapid change in temperature could break the glass rod or dynamo or LED parts.
12

Publicidad

Tabla de contenido
loading

Tabla de contenido