Hardware ReferenceÑFlextight Precision II
20
Calibrating the Scanner
2.
Load the calibration target into the 6X6 original holder with the
stripes pointing in towards the scanner (vertically). Take care to
place the target as straight as possible. Use the scannerÕs light table
grid and the corners marked on the target to help align it.
3.
Select Calibration from the Maintenance menu.
4.
Click on the Transmission button. A window appears in which you
can choose the resolutions you want to calibrate.
5.
Choose which resolutions you want to calibrate. By default, all res-
olutions are marked and will therefore be calibrated. To prevent
one or more resolutions from being calibrated, click on the appro-
priate box to remove the mark. We recommend that you calibrate
for all resolutions.
6.
Click on the Calibrate button. The scanner will focus for each of the
selected resolutions and calibrate for each of the selected resolu-
tions (one at a time).
7.
When the calibration is Þnished, ColorFlex will again display the
window in which you can select the calibration type.
8.
At the scanner, replace the transmission target with the reßex tar-
get, which is a white piece of paper with many black lines on it.
Position it just as you did the reßex target.
Imacon Flextight Precision II UserÕs Guide