NOTE:..Your.computer.may.have.other.units.of.measure.
programmed.into.it..If.so,.holding.the.CALIBRATE.
button.and.momentarily.pressing.the.DISPLAY.button.
will.toggle.through.all.factory.set.units..Other.pos-
sible.units.are:.IGL.(imperial.gallon),.QT.(quart),.CF.
(cubic.feet),.CM.(cubic.meter),.BL.(42.gal..barrel),.
CC.(cubic.centimeter).or.OZ.(ounce).
Switching. between. different. units. will. not. corrupt. the.
Total's.contents..For.example,.in.GL.mode,.the.computer.
totalizes.10.00.gallons,.if.the.user.switches.to.LT.mode,.
the.display.will.read.37.85.litres.(the.same.volume,.dif-
ferent.unit).
The."field".calibration.may.be.set.by.the.user,.and.can.
be.changed.or.modified.at.any.time.using.the.calibration.
procedure.described.in.the.Calibration.Section..Totals.
derived.from.the.field.calibration.are.invoked.when.the.
FAC.icon.is.no.longer.visible.on.the.top.line.of.the.display.
CALIBRATION
Field Calibration Procedures
(K-Factor Entry Method)
If.the.display.was.purchased.as.an.accessory.or.remote,.
then. the. field. calibration. method. set. at. the. factory. is.
K-factor.Entry.method.
This. method. allows. the. user. to. key. in. a. single. point.
K-factor.value.that.represents.the.meter.it.will.be.used.
with..K-factor.values.for.specific.meters.can.be.found.
on.the.meter.itself,.in.Table.1.or.www.gpimeters.net.
1.. To.field.calibrate,.press.and.hold.both.CALIBRATE.
and.DISPLAY.buttons.for.about.3.seconds.until.you.
see.FLdCAL..Release.both.buttons.and.you.will.see.
Kxxxx.x.(where."x".represents.the.current.field-cal.
k-factor.value)..You.are.now.in.the.field.calibration.
mode.
2.. The.far.left.digit.will.be.blinking..The.DISPLAY.button.
can.then.be.pressed.to.select.the.digit.location.and.
the.CALIBRATE.button.can.be.pressed.to.scroll.the.
desired.value.at.the.blinking.position..Edit.the.K-factor.
shown.to.the.desired.value.. A cceptable.K-factor.range.
is.0000.1.to.9999.9.
3.. After.the.new.value.has.been.entered,.momentarily.
press.and.release.both.buttons.."CALEND".will.be.
momentarily.displayed..Unit.is.now.ready.for.use.
4.. Notice.that.the.upper.display.line,.the."FAC".icon.and.
all.the.units.of.measure.have.disappeared.
Alternate.units.of.measure.are.not.selected.when.the.
meter.is.operating.with.field.calibration..This.calibration.
is.a.unique.single-point.calibration.for.the.meter.and/
or.application.
NOTE:.To.return.to.factory.calibration.(FAC),.press.and.
hold.both.CALIBRATE.and.DISPLAY.buttons.for.about.
3.seconds,.until.FAcCAL.is.displayed..Then.release.
buttons..Unit.should.return.to.normal.operation.and.
FAC.icon.is.visible.
6
NOTE:.If.the.field.calibration.mode.is.entered.and.NO.
fluid.is.dispensed,.then.upon.leaving,.the.computer.
will.use.data.from.the.last.successful.field.calibration.
T A B L E 1
Model
Size
. G2_05.
1/2.in..
. G2_07.
3/4.in..
. G2_10.
1.in..
. G2_15.
1-1/2.in..
. G2_20.
2.in..
. G2P05.
1/2.in..
. G2P10.
1.in..
. G _T-051.
1/2.in..
. G _T-075.
3/4.in..
. G _T-100.
1.in..
. G _T-150.
1-1/2.in..
. G _T-200.
2.in..
. G _T-300.
3.in..
. G _P-050.
1/2.in..
. G _P-051.
1/2.in..
. G _P-075.
3/4.in..
. G _P-100.
1.in..
. G _P-150.
1-1/2.in..
. G _P-200.
2.in..
.GM001.
1/8.in..
.GM002.
1/4.in..
.GM003.
1/4.in..
.GM005.
1/2.in..
.GM006.
3/4.in..
.GM007.
1.in..
.GM010.
1.in..
.GM015.
1-1/2.in..
.GM020.
2.in..
Field Calibration Procedures
(Correction Factor Method)
If.the.display.was.purchased.with.a.meter.body.then.the.
field.calibration.method.set.at.the.factory.is.the.Correc-
tion.Factor.method.
This.method.allows.the.user.to.tweak.the.factory.cali-
bration.by.a.percent.that.represents.application,.fluid.or.
plumbing.differences.
1.. To. calibrate,. press. and. hold. the. CALIBRATE. and.
DISPLAY.buttons.for.about.3.seconds.until.you.see.
FLdCAL.. Release. both. buttons. and. you. will. see.
CF-00.0..You.are.now.in.the.field.calibration.mode.
and.values.from.-99.9%.to.+99.9%.can.be.entered.
Typical K-Factor
(pulses / gallon )
2500
1100
560
215
100
2400
540
10000
3750
896
350
181
50
10000
10000
3750
896
340
181
5855
3785
1514
424
197
197
136
55
25