SAVE THESE INSTRUCTIONS
Operations Guide for
GPI 4-20 mA Out with Display
GX510
TABLE OF CONTENTS
English ........................................................................1
Spanish.....................................................................18
E N G L I S H
GENERAL INFORMATION
This manual will assist you in operating and maintain-
ing the GPI electronics supplied with your GPI meter or
as an accessory unit on both local and remote models.
The GPI electronics can be used in indoor or outdoor
applications where occasional exposure to moisture is
common.
Display information in this manual supersedes display information provided with your meter.
04/12
Local Models
GM Series
(Shown here on GM001)
www.gpimeters.net
1-888-996-3837
Remote Model
The 4-20 mA Out with Display is available in two
versions. One indicates flowrate in units/minute and
one in units/hour. Both indicate flow totals in gallons
and litres.
The 4-20 mA Out with Display can be used on all
GPI models, including the Precision G series, the
Industrial Grade G2 series, the Commercial Grade
A1 series and the Positive Displacement GM series
oval gear meters.
Product differences in this manual are identified by
either, Local or Remote as necessary.
If the meter was purchased with this display, then it will
come calibrated from the factory for gallons "GAL" and
litre "LTR". Field calibration is also available.
If the unit was purchased as an accessory or remote,
the calibration has not been entered and the end user
will need to configure and calibrate the display.
5252 East 36th Street North
Wichita, KS USA 67220-3205
TEL: 316-686-7361
FAX: 316-686-6746
GX500
1
Rev. - 920887-01

Publicidad

Tabla de contenido

Solución de problemas

loading

Resumen de contenidos para GPI GM Serie

  • Página 1 This manual will assist you in operating and maintain- If the unit was purchased as an accessory or remote, ing the GPI electronics supplied with your GPI meter or the calibration has not been entered and the end user as an accessory unit on both local and remote models.
  • Página 2: Safety Instructions

    The GPI 4-20 mA Out with Display is a flow totalizer The following installation guidelines are separated by and rate meter with industry standard current loop out- meter series and mounting type. put. The unit is loop powered, and provides a 4-20 mA...
  • Página 3: Connecting The Equipment

    If you require a longer cable, a 100 ft. cable kit is EDM PWR: Local Vcc. Regu- LOOP (–) available from GPI, or use Belden 9363 cable. When lated 5-VDC internal power. wiring longer lengths of cable, be sure to connect...
  • Página 4 Customer Equipment With Built-in Power Supply – High Temp Applications Input: Standard Remote Sensor (Variable Reluctance Pickup Coil) Output: Customer Equipment, 0-20 mA Sensing, Built-in Loop Power Supply INPUTS OUTPUTS Loop (+) GPI Meter 4-20 mA Out (current) with Display Unpowered Coil-A...
  • Página 5 W I R I N G D I A G R A M 3 — 4-20 mA or 0-20 mA Output — Customer Equipment Without Built-in Power Supply Input: Turbine Mounted Display or Conditioned Signal Sensor (See inputs from Diagram 1) Standard Remote Sensor (See inputs from Diagram 2) Output: Customer Equipment, 0-20 mA Sensing, Separate Power Supply Loop Power Supply INPUTS...
  • Página 6 W I R I N G D I A G R A M 5 — Pulse Output — Customer Equipment Without Built-in Power Supply Input: Turbine Mounted Display or Conditioned Signal Sensor (See inputs from Diagram 1) Standard Remote Sensor (See inputs from Diagram 2) Output: Customer Equipment, Frequency Sensing, Separate Loop Power Supply Loop Power Supply INPUTS...
  • Página 7 W I R I N G D I A G R A M 7 — 4-20 mA or 0-20 mA Output — Customer Equipment With Built-in Power Supply Input: Hall Effect* Output: Customer Equipment, 0-20 mA Sensing, Built-in Loop Power Supply HE Power Supply 4.5 - 24 VDC INPUTS...
  • Página 8 W I R I N G D I A G R A M 9 — 0-5 V Output — Customer Equipment Without Built-in Power Supply Input: Reed Switch (See inputs from Diagram 6) or Hall Effect* (See inputs from Diagram 7) Output: Customer Equipment, 0-5 V Sensing, Separate Loop Power Supply Loop Power Supply INPUTS...
  • Página 9: Configuration

    NOTE: Your computer may have other units of measure Access to the configuration settings require a specific programmed into it. If so, holding the CALIBRATE procedure and a pin code available through the GPI button and momentarily pressing the DISPLAY button website at www.gpimeters.net or call GPI Customer will toggle through all factory set units.
  • Página 10 1. To field calibrate, press and hold both CALIBRATE NOTE: To return to factory calibration (FAC), press and and DISPLAY buttons for about 3 seconds until you hold both CALIBRATE and DISPLAY buttons for about see FLdCAL. Release both buttons and you will see 3 seconds, until FAcCAL is displayed.
  • Página 11: Lockout Feature

    NOTE: A true loop current of “0” in a loop powered de- the procedure, it should be registering at or very vice like the GPI transmitter is not obtainable. That’s near 20 mA (within the resolution specifications for because the current loop powers the transmitter, the present conditions).
  • Página 12 • Longer (slower) response times are also appropriate in situations where sensor-output frequency fluctu- ates or wobbles substantially. The GPI 4-20 mA Out with Display offers a choice of five response-time settings, selectable by the unit’s pushbuttons. Procedure 1. Start with the unit unpowered. If the unit is presently operating, temporarily disable its external power supply.
  • Página 13: Troubleshooting

    TROUBLESHOOTING SYMPTOM PROBABLE CAUSE CORRECTIVE ACTION A. METER IS NOT 1. Field Calibration not performed Field Calibrate again or select Factory Calibration. ACCURATE properly 2. Factory Calibration not suitable Perform a Field Calibration according to Calibration Section or for liquid being measured select the proper Factory Calibration selection (i.e., gallon or litre).
  • Página 14: Dimension Drawings

    DIMENSION DRAWINGS...
  • Página 15 ILLUSTRATED PARTS DRAWING Item Item Part No. Description Req’d. Part No. Description Req’d. 120512-01 Switch Keypad Kit .......... 1 904005-13 Screw, 6-32 x 1/2 in........4 120048-01 Gasket ............1 904002-44 Screw, 8-32 x 5/16 in........2 12051803 Computer Kit (Hours) ........
  • Página 16 Batch Total: SPECIFICATIONS – LOCAL MODEL Up to 999,999 (x100) Applications: Cumulative Total: Use for indoor or outdoor applications where occa- Up to 999,999 (x100) sional moisture is common. Temperature: Materials: Ambient Temperature: +32° F to 140° F (0° C to 60° C) Acetal, Amorphous Nylon, Silicone Rubber, Polyes- Cable: ter (decals), FKM (gasket &...
  • Página 17 1.1 lb. (.5 kg) Digital (Pulse Out): Open Collector, square wave. Will To power some low power pickup/display devices, such as a GPI switch up to 60 V and up to 200 mA. Closed circuit “EDM” module, a loop powered GX500 can supply 5 VDC at up voltage drop typically 0.1 V;...
  • Página 18: Weee Directive

    8:00 a.m. to 5:00 p.m. Central time. imiento de la electrónica de GPI que se suministran Tel: 316-686-7361 con su medidor de GPI, o como una unidad de acceso- Fax: 316-686-6746 rio en los modelos locales y remotos. La electrónica...
  • Página 19: Instrucciones De Seguridad

    • Cuando se aplica energía externa al transmisor, utilice corriente continua. Instale el GPI de salida de 4-20 mA con pantalla usan- do los pernos, tornillos o pernos estándar U para tu- • Desconecte la alimentación externa para el trans- berías.
  • Página 20: Configuración

    • Si necesita un cable más largo, un kit de 100 pies de cable está disponible a través de GPI, o usar Las características disponibles incluyen: Belden 9363 por cable. Cuando la longitud de ca- • Totales 0 a 3, ya sea reajustable (por lotes) o que bleado de cable más largos, asegúrese de conec-...
  • Página 21: Caudal De Funciones

    Cuando el total acumulativo alcanza una lectura de la CALIBRACIÓN pantalla de 999,999 el equipo seleccione un icono de X10. Esto indica al operador que un cero debe ser aña- Los procedimientos de calibración de dido a los 6 dígitos mostrados. Cuando la reinversión campo (Método de Factor K) se produzca la próxima, el equipo se seleccione un icono de X100.
  • Página 22: Configuración De 4-20 Ma Extremos

    Ajuste de la baja (4 mA) Punto 1. Para calibrar, presione y mantenga presionados los botones CALIBRATE y DISPLAY durante unos extremo final: 3 segundos hasta que vea FLdCAL. Suelte los Para configurar de 4 mA a flujo cero, vaya al paso 3. botones y verá...
  • Página 23: Función De Bloqueo

    • El funcionamiento de un medidor de caudal en la de GPI no se puede obtener. Eso es porque los alta pueden requerir un menor (más rápido) tiempo poderes de bucle de corriente del transmisor, y su...
  • Página 24: Mantenimiento

    El 4-20 mA de GPI con dispositivo visual ofrece una SOLUCIÓN DE PROBLEMAS selección de cinco de tiempo de respuesta ajustables, seleccionables mediante botones de la unidad. A. EL CONTADOR NO ES EXACTO. 1. La calibración de campo no se ha realizado Procedimiento correctamente.
  • Página 25: Especificaciones - Local

    F. LA SALIDA DEL LAZO ES ESTABLE, PERO EL ESPECIFICACIONES - LOCAL TIEMPO DE RESPUESTA ES MUY LENTO. 1. La respuesta de salida de ajuste de tiempo Aplicaciones: es demasiado larga, especialmente para la Se utiliza para aplicaciones en interiores o al aire señal de entrada rápida.
  • Página 26: Conexiones Eléctricas

    Velocidad de Respuesta: Después del cambio de ESPECIFICACIONES - REMOTO paso en la frecuencia de entrada, de salida del cir- cuito garantizado estable dentro de la acumulación Aplicaciones: de tiempo de 3 x (acumulando vez que el usuario Se utiliza para aplicaciones en interiores o al aire puede seleccionar de 0,3 segundos, 0,7 segundos, libre donde la humedad ocasional es común.
  • Página 27: Temperatura Ambiente

    Servicio de Atención al Cli- Junto medidor de flujo puede añadir errores adicio- ente de GPI en Wichita, Kansas, de lunes a viernes, de nales. 8:00 am a 5:00 pm hora Central.
  • Página 28 To make a claim against this warranty, contact the GPI Customer Service Department at 316-686-7361 or 888-996-3837. Or by mail at: Great Plains Industries, Inc.