Descargar Imprimir esta página

Connecting The Power Supply - RPB GX4 08-400 Manual Del Usuario

Ocultar thumbs Ver también para GX4 08-400:

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

CONNECT THE AIR SUPPLY
FIGURE 2.1
Insert the 3/8" O.D. Air Supply Hose (P/N:
08-428) into the push-lock fitting on the left
side of the monitor. To disconnect, press the
outside ring of the push-lock and pull the
hose out. Note: Hose can be trimmed to the
desired length.

CONNECTING THE POWER SUPPLY

FIGURE 3.1
Connect the power supply by screwing it onto
the receptacle on the left side of the unit (next
to the air inlet fitting.) Plug the AC Adaptor
into a power source.
FIGURE 2.2
Connect the air supply hose to a Grade D
air source regulated to between 7 to 80 PSI
(50-550 kPa). Maximum air flow is achieved
at 20 PSI. It is recommended to supply the
with 15-20 PSI. Max pressure is 80
GX4
®
PSI. Use the RPB
Regulator Assembly
®
08-470 to regulate the air pressure.
FIGURE 3.2
1
+
When using Battery Clips (08-431), connect:
Black = Negative
Red = +12V (± 0.5V)
between 10-125psig.
2
-
9