Cleaning and Disinfecting Your Meter and Lancing Device
1. Wash hands with soap and water and dry thoroughly.
2. Inspect for blood, debris, dust, or lint anywhere on the
meter or lancing device.
3. To clean the meter, use a moist (not wet) lint-free cloth dampened with
a mild detergent. Wipe all external areas of the meter or
lancing device including both front and back surfaces until visibly
clean. Avoid wetting the meter test strip port.
NOTE:
• Do not get water inside the meter or immerse the meter or lancing
device in water or any other liquids.
• Do not use glass or household cleaners on the meter or the lancing
device.
• Avoid using ammonia to clean the meter or lancing device.
• Avoid wetting the test strip port of the meter.
4. To disinfect your meter, clean the meter with one of the validated
disinfecting wipes listed below. Other EPA registered wipes may be
used for disinfecting the EVENCARE G2 system, however these other
wipes have not been validated and could affect the performance of
your meter.
• Dispatch
Hospital Cleaner Disinfectant Towels with Bleach
®
(EPA Registration Number: 56392-8)
• Medline Micro-Kill+™ Disinfecting, Deodorizing, Cleaning Wipes with
Alcohol (EPA Registration Number: 59894-10)
• Clorox Healthcare
(EPA Registration Number: 67619-12)
• Medline Micro-Kill™ Bleach Germicidal Bleach Wipes
(EPA Registration Number: 69687-1)
46
Caring for the Meter
Bleach Germicidal and Disinfectant Wipes
®