SETTINGS (continued)
Emissivity and Compensation
Emissivity
Enter the emissivity of the target. Target emissivity can be determined
Setting
experimentally, or estimated using an emissivity table. For more information,
contact Calex. Minimum: 0.2. Maximum: 1.0.
Enable
If enabled, compensates for errors caused by reflected energy from hotter or
Reflected
colder objects.
Energy
Compensation
Reflected
Enter the temperature of the surroundings of the target for Reflected Energy
Temperature
Compensation. Minimum: -20°C. Maximum: 1000°C.
4 to 20 mA Output (-CRT models)
Set the temperature range limits for the 4 to 20 mA output.
Temperature
The lower temperature range limit. Minimum: -20°C. Maximum: 900°C.
at 4 mA
Temperature
The upper temperature range limit. Minimum: 80°C. Maximum: 1000°C.
at 20 mA
Please note
The difference between the temperatures at 4 mA and at 20 mA must be at
least 100°C. The temperature at 20 mA must be greater than the temperature
at 4 mA.
Modbus Address (-BRT models)
Modbus
The current Modbus address of the sensor is displayed. Enter a new address,
Address
then press Apply to save it to the sensor. Cycle the power to use the new
address.
Minimum: 1. Maximum: 247.
Alarms
Configure the settings for Alarm 1 and Alarm 2 separately, and configure
alarm logging settings.
Manually Reset Alarms
If an alarm has been triggered, allows both alarms to be triggered again.
Alarms will not be triggered again until they are reset, either automatically or
manually.