4.0 ERROR CODES
During the initial power-on testing or during operation, the scale may show an
error message. If an error message is shown, repeat the step that caused the
message. If the error is still shown contact your dealer for further support.
ERROR
Err 1
Err 2
Err 4
Err 6
© Adam Equipment Company 2011
DESCRIPTION
Time input error.
Date input error
Initial Zero is greater than
allowed (typically 4% of
the maximum capacity)
when power is turned on
or when [Zero] is
pressed,
A/D count is not correct
when turning the scale
on.
POSSIBLE CAUSES
Tried to set an illegal time, i.e.
26 hours
Tried to set an illegal date, i.e.
th
36
day
Weight is on the pan when
turning the scale on.
Excessive weight on the pan
when zeroing scale.
Improper calibration of the
scale.
Damaged load cell.
Damaged Electronics.
Platform is not installed.
Load cell is damaged.
Electronics is damaged.