NOMENCLATURE

Hydrocollator ® Heating Units
WHITE LAMP (where applicable), when illuminated, indicates that main power is
connected, the switch is on and unit is operating.
RED LAMP (where applicable), when illuminated, warns of an overheat/low water
condition. Action is necessary to correct the condition.
GREEN LAMP (220/240 volt units), when illuminated, indicates that main power is
connected, the switch is on and unit is operating.
White Lamp
IMPORTANT: Some models are fitted with a thermal cut-out thermostat which acts
as a "non self-resetting" thermal cut-out. In the unlikely event of a fault condition, this
device is designed to open the power circuit between 88° C ± 8° C (190° F ± 14° F). It is
not possible to alter its operating range. The operation of the unit may be restored by
depressing the red button after the water/element temperatures have been allowed to
cool. If the fault condition persists, please contact DJO, LLC or an authorized agent.