a) Verify if the Injector is in the charged or uncharged state by confirming
the position of the Injector Hammer as shown in figures 1a (uncharged)
and 1b (charged). If the Injector is charged and the Handles are closed,
proceed to Section 2) Fill Syringe. DO NOT open the Handles when the
Injector is charged.
b) When the Injector is in the uncharged state, as shown in figure 1a, and
the Handles are closed, open the Handles to prepare the Injector. Press
the rear Injector Latch Button to allow the Injector Handles to open
freely, as shown in figure 2.
NOTE: If the Injector is in the charged state, as shown in figure 1b, and
the Handles are partially open, as shown in figure 3, close the Handles and
proceed to Section 2) Fill Syringe. DO NOT touch internal components or
attempt to fully open the Handles when the Injector is charged.
Workflow
1) Prepare Injector
figure 1a
figure 2
figure 3
figure 1b