Guidance and Manufacturer's Declaration – Electromagnetic Emissions
The BabyCO is intended for use in the electromagnetic environment specified below.
The customer or the user of the BabyCO should assure that it is used in such an
environment
Emission Test
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC61000-3-2
Voltage fluctuatuins /
flicker emissions
IEC61000-3-3
Guidance and Manufacturer's Declaration – Electromagnetic Immunity
The BabyCO is intended for use in the electromagnetic environment specified below.
The customer or the user of the BabyCO should assure that it is used in such an
environment
Immunity Test
Electrostatic
discharge (ESD)
IEC61000-4-2
Electrical fast
transient / burst
IEC61000-4-4
Surge
IEC61000-4-5
Voltage dips,
short
interruptions
Compliance
Group 1
Group B
Not Applicable
(battery powered)
Not Applicable
(battery powered)
IEC 60601 test
level
± 6 kV contact
± 8 kV air
± 2 kV for power
supply lines
± 1 kV for input /
output lines
± 1 kV line(s) to
line(s)
± 2 kV line(s) to
earth
<5% U
T
(>95 % dip in U
T)
For 0.5 cycle
Electromagnetic Enviroment -
Guidance
The BabyCO uses RF energy only for its
internal function. Therefore, its RF
emissions are very low and are not likely
to cause any interference in nearby
electronic equipment
The BabyCO is suitable for use in all
establishments, including domestic
establishments and those directly
connected to the public low-voltage
power supply network that supplies
buildings used for domestic purposes
Compliance
Electromagnetic
Level
Enviroment -Guidance
± 6 kV contact
Floors should be wood,
± 8 kV air
concrete or ceramic tile. If
floors are covered with
synthetic material, the
relative humidity should be
at least 30%.
Not Applicable
Mains power quality should
(Battery
be that of a typical
Powered and
commercial or hospital
Serial cable
environment.
<3m)
Not Applicable
Mains power quality should
(battery
be that of a typical
Powered)
commercial or hospital
environment.
Not Applicable
Mains power quality should
(battery
be that of a typical
Powered)
commercial or hospital
29