TECHNICAL DATA
Materials :
 AISI 316 and PVC Body
 Special Glass Optics
 NBR and Sylicon O-Rings
Thread: 3/4" BSP
Measuring ranges:0,00 mg/L to 20,00 mg/L
Measuring method: Optical measure by luminescence
Calibration method:
Calibration on air: 1-point calibration, Water
-
saturated with air at 100%
Calibration through sample: comparison with a
-
standard tool
Calibration on 0% solution: 1-point calibration, specific
-
solution at 0% of oxygen saturation
Accuracy:
± 0,2 mg/L when < 5mg/L
± 0,3 mg/L when > 5mg/L
Resolution: 0,01 mg/L
Repeatability: ± 0,1 mg/L
Response: T
< 60s
90
Maximum refreshing time: < 1 second
WorkingTemperature: 0÷50 ° C Accuracy ± 0.2°C
Max Working Pressure: 5 bar
Maximum absorption: 2W
Mechanical Protection: IP68 Sensor+cable
Cable: 10m integral
Power Supply: 12...24Vdc
Signal interface: RS-485 Modbus RTU Protocol (4-20mA opt.)
Water move: No necessary move
Compensation of temperature: Via internal NTC
Luminophore diameter: 10mm
Connector IP67: yes
Order codes
9700830067
9700831067
9700832067
9700833067
9400410134
9400410135
S423/C/OPT Optical Oxygen Probe – Data Sheet
S423/C/OPT Optical Oxygen and Temperature Probe
S423/C/OPT PVC Optical Oxygen and Temperature Probe
S423/C/OPT 4-20mA Optical Oxygen and Temperature Probe
S423/C/OPT 4-20mA PVC Optical Oxygen and Temperature Probe
Luminescent membrane with AISI316 support for S423/C/OPT
Luminescent membrane with PVC support for S423/C/OPT
S
4
2
3
/
C
/
O
P
S
4
2
3
/
C
/
O
P
O
x
y
g
e
n
O
x
y
g
e
n
T
O
p
t
i
c
a
l
T
O
p
t
i
c
a
l
P
r
o
b
e
P
r
o
b
e
D
a
t
a
S
h
e
D
a
t
a
S
h
e
DIMENSIONS
e
t
e
t
15

Publicidad

Tabla de contenido
loading

Tabla de contenido