Installation
Step 1: Mount eyewash pedestal to wall (optional)
CAUTION:
Eyewash station weighs 60 lbs when filled to capacity.
Assistance from another individual may be required to
safely install the unit. Mounting bracket screws and wall
anchors must be strong enough to support the filled unit.
1. To mount the pedestal to the wall, follow the procedure
outlined below:
• Secure the mounting bracket to the wall with screws and
wall anchors (provided by installer).
2. Place the pedestal into the heater jacket. Insert the tube from
the drain hole at the bottom of the pedestal through the hole in
the back of the heater jacket. Drain hole location is shown on
page 5.
3. Attach the pedestal to the mounting bracket at the handles on
the back of the unit.
4. Check condition of gasket; ensure clamps are out of the way.
Check location of heater pad and routing of its power cable.
Ensure area is clear to receive water tank.
When mounting the eyewash unit to the mounting bracket, make
sure the unit is level and the sprayheads are at a height of 33-45"
from the floor and situated at least 6" from the wall or nearest
obstruction (ANSI Z358.1-2009).
Step 2: Fill and mount tank
1. Pour clean, potable, tepid water into the tank; continue pouring
until the tank is 1/3 to ½ full.
2. Add water preservative (not included) to the eyewash tank.
Follow the manufacturer's instructions for the proper use of
water preservative.
3. Add clean, potable, tepid water to the tank until the water level
reaches the fill line.
4. Insert one liner into the cap (with hole) and attach the cap to
the top of the tank.
• Check location of heating pad and thermostat on tank as
shown in the illustration.
FOR TANK INSTALLATION: To ensure proper installation, the tank
should be placed on the pedestal from the front of the unit.
5. Invert the tank and place it on top of the pedestal as shown on
page 5. The pierce on the pedestal will puncture a hole in the
cap liner and allow solution to flow.
6. Raise the arm to snap fit in an upright position. A slight
adjustment to the position of the tank may be required for
proper level placement of the tank on the pedestal.
• Check to make sure jacket is not pinched under the tank and for proper routing of all cables.
7. Secure the tank to the pedestal using the side tank clamps. Make sure the latches are pulled down tight to the
pedestal (see page 5).
8. Lower the arm to activate the water flow. Allow the water to flow until air bubbles form in the tank. Raise the arm and
"snap" in the upright position. The system is now ready for use.
9. Insert the 5-pin connector from the tank heater assembly into the mating socket on the junction box and screw tight.
Slide the heater jacket up to cover the tank and secure the jacket with the hook-and-loop fasteners. Slide the arm into
pocket of the jacket flap.
After the tank is mounted to the pedestal, minor spillage may occur. Allow the fluid to stabilize completely before inspecting the unit for
leaks.
4
8/9/2012
33"
(838mm)
to 45"
(1143mm)
Cap with Hole
Liner (insert into cap
before attaching cap