Cleaning the Nebulizer After Each Use
WARNING: To prevent possible risk of infection from contaminated
medication, clean the nebulizer as follows after each treatment.
1.
Ensure the plug is disconnected from the device's power socket.
2.
Remove the top cover, nebulization chamber, air-flow chamber, and
medication cup from the device. Remove the filter from the cover.
3.
Completely empty the conduction liquid in the reservoir.
4.
Wipe the main unit with a clean, soft, moistened cloth.
5.
Wash the top cover, nebulization chamber, air-flow chamber,
mouthpiece, masks, and medication cup in warm, soapy water, and
rinse thoroughly. Allow to air dry. Store in a clean, dry location.
Cleaning the Nebulizer Daily
WARNING: To prevent possible risk of infection from contaminated
medication, clean the nebulizer as follows daily.
Once daily, after washing as above, soak the washed components in a
fresh solution consisting of one part warm water and one part white
vinegar for 30 minutes followed by a warm water rinse. Allow to air dry.
Store in a clean, dry location.
Replacing the Medication Cup
Replace the medication cup after every two treatments. There is a
package of 20 replacement medication cups supplied with the nebulizer.
Additional medication cups can be purchased from Graham-Field.
14
20 Replacement
Medication Cups
6700-INS-LAB-RevC14