Blue Connect Manual Del Usaurio página 7

Ocultar thumbs Ver también para Connect:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 24
Blue Connect maintenance
A. Calibration
NB: Calibration is not needed when receiving the device. Usually, calibration is needed once or twice
a season. Use the specially designed calibration kit for Blue Connect or standard calibration solutions
available in your pool store. Calibration products are for single use only. Never use them twice.
1.
Take out the device from the water, pull out the usage cap and rinse the probe in clear water.
2.
Rinse the storage cap with clear water and wipe it.
3.
Activate the Internet (Wi-Fi or 4G) and Bluetooth® on your smartphone.
NB: Stay close to your device throughout the whole calibration process.
4.
In the app, in "Settings" » "My Blue Connect" » "Probe settings", press "Calibrate Blue
Connect ". If you can choose among several products, select the calibration solution you
will use.
5.
Place the calibration solution in the storage cap and screw it to the device. Then, press
"Next". At the top of the screen, you will see the progression of the measurement. Once the
measurement is complete, unscrew the storage cap and rinse the probe. Empty, rinse
wipe the cap.
Always clean the probe and storage cap with clear water after using a calibration product to avoid mixtures.
7
7
and
7

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

Fit50

Tabla de contenido