Dosper evo
11. Replacement of the fuse
Disconnect the power cord from the
current supply.
The fuse holder is integrated in the
mains connection on the back of the
machine.
For the control or the change of the fuse
press the upper and the lower latch of
the fuse holder simultaneously and take
the holder out.
34
Pull the defective fuse out of the clamp
holder by pulling gently. Replace with
a new, identical fuse.
Important notice: Only use fuses with
protection power (T 1, 6A) as stated on
the back of the machine.
After the change insert the fuse holder
again by pressing gently until the fixa-
tion locks noticeably.