Troubleshooting Guide

Problem
Illumination spot is too high
and is out of field of view.
User encounters glare off of the
pneumatic insufflation system.
18
Solution
Make sure you are 14" to 16" from
patient or object under inspection.
If you are too close, the illumination
spot will be too high.
Make sure optical section is aligned
with user's eyes. Sug-gest looking into
the instrument when on user. The user's
eyes should be centered vertically on
the center mirrors.
A. optical section
too high on user
B. optical section
too low on user
C. optical section
centered correctly
on user
To correct A and B, simply loosen screw
on back of optical section and adjust
optical section up or down until image
is centered.
Clean the glass section with alcohol
swab or clean cotton cloth.
Rotate pneumatic assembly
until glare dissipates.
Problem
Instrument does not illuminate.
Image is distorted or double
image is seen.
How do I utilize bifocals?
Solution
Replace 08500 lamp.
See page 11 for instructions.
Battery is expired or discharged.
See page 15 and 16.
Instrument must be set correctly for
interpupillary distance adjustment.
See page 9 for instructions.
Depth perception is slightly altered
with 3D optical design. Continued
usage will improve comfort level.
Install bifocal lenses so that only the
upper third is for distance and lower
two-thirds is for reading. Adjust the
nose bridge so that frame sits higher
on nose. Then lower the height
adjustment. This will ensure that
the binocular viewing system will
be positioned in front of the reading
part of the lenses.