General information regarding the CE marking
HAMEG instruments fulfill the regulations of the EMC directive. The conformity test
made by HAMEG is based on the actual generic- and product standards. In cases
where different limit values are applicable, HAMEG applies the severer standard. For
emission the limits for residential, commercial and light industry are applied. Regarding
the immunity (susceptibility) the limits for industrial environment have been used.
The measuring- and data lines of the instrument have much influence on emmission
and immunity and therefore on meeting the acceptance limits. For different applications
the lines and/or cables used may be different. For measurement operation the following
hints and conditions regarding emission and immunity should be observed:
1. Data cables
For the connection between instruments resp. their interfaces and external devices,
(computer, printer etc.) sufficiently screened cables must be used. Without a special
instruction in the manual for a reduced cable length, the maximum cable length of a
dataline must be less than 3 meters. If an interface has several connectors only one
connector must have a connection to a cable.
Basically interconnections must have double screening. For IEEE-bus purposes the
double screened cables HZ72S and HZ72L from HAMEG are suitable.
2. Signal cables
Basically test leads for signal interconnection between test point and instrument should
be as short as possible. Without instruction in the manual for a shorter length, signal
lines must be less than 3 meters long.
Signal lines must screened (coaxial cable - RG58/U). A proper ground connection is
required. In combination with signal generators double screened cables (RG223/U,
RG214/U) must be used.
3. Influence on measuring instruments.
Under the presence of strong high frequency electric or magnetic fields, even with
careful setup of the measuring equipment an influence of such signals is unavoidable.
This will not cause damage or put the instrument out of operation. Small deviations of
the measuring value (reading) exceeding the instruments specifications may result
from such conditions in individual cases.
HAMEG GmbH
Subject to change without notice