Step 4
I
Use your DermapenCryo™
Pressing
the
black
trigger
DermapenCryo™ will initiate the flow of
cryogenic liquid. To stop, just put the cap
back on DermapenCryo™, gently squeeze
the trigger switch to start the working.
Releasing the trigger will close off the
flow of cryogenic liquid. You may observe
that during 1 or 2 seconds after releasing
the trigger, the remaining gas in the
applicator is discarded.
on
you
Blue dot applicator
for applications of 1-3 mm
White dot applicator
for applications of 2-5 mm
I m p o r t a n t n o t i c e
• Always gently manipulate all components of DermapenCryo™
instruments. Never use more than gentle hand force. Never use
pliers or other tools to manipulate the instrument.
• Incorrect placement of the filter/opener device into the loading
cavity will cause irreversible damage to the instrument.
• Residual foreign matters in the loading cavity while placing a new
filter/opener can cause instrument failure.
• When loaded, the instrument is under high pressure. Any change to
the construction of the instrument may cause a potential risk.
• Never block the gas release opening at the end of DermapenCryo™'s
backend!
• All external parts can be wiped with a cloth soaked in any
noncorrosive disinfecting solution. Never immerse the entire
instrument in a disinfecting solution. Only the applicators can be
autoclaved.
• The applicators and the cartridges must be kept away from dust
and high temperatures. Leave the cap on when not in use.
• The DermapenCryo™ should be handled with care. When storing
the instrument with a loaded cartridge in its case, position the
DermapenCryo™ in such a way that the trigger switch cannot
be depressed when closing the lid of the case. This would cause
undesired gas flow.
• DermapenCryo™ instruments are intended for professional use only.
• Best results are obtained with instrument and cartridges at room
temperature (21°C /77°F). Lower temperatures will give less pressure
in gas flow with longer flow time. Higher temperatures will cause
higher gas pressure and shorter flow time.
• Empty cartridges should be discarded as metal scrap.