VI.3 - Lubrication
E
Follow the spray lubricant manufacturer's instructions.
Wear appropriate protective clothing (splashes, etc.).
Keep away from all sources of heat or ignition. In particular: do
not smoke. Risk of inflammation.
After cleaning and before each sterilisation (once a day minimum):
• Remove the rotary instrument,
• Insert the tip to the back of the MD (Fig 6),
• Cover the head of the MD with a soft cloth or paper or a wipe,
• Point the head towards the bottom,
• Spray several times by firmly holding the instrument,
• Wipe excess oil with a cloth or a wipe.
I.4 – Operational test
• Connect the MD to a micro-motor, point the head towards the
bottom.
• Turn the micro-motor on at low speed for 30s. Gradually bring
up to maximum speed.
• Wipe excess oil with a cloth or a wipe
If you notice overheating, irregularities, vibrations, abnormal
noises when using the MD, immediately contact your AS.
VI.5 - Sterilisation
g
• The instruments are to be sterilised before first use and
after each use.
• Only sterilise cleaned, lubricated and tested instruments.
• We recommend sterilisation that complies with the EN 13060
standard, class B.
• Independently put each MD in a sterilisation sachet that
complies with current standards and the size of the MD.
• Comply with the space between the sachets and do not overload
the autoclave.
• These MDs and their accessories (apart from accessories
x
with the logo
at 135°C for 18 minutes minimum (sterilisation time).
x
Check that the device is completely dry at the end of the
cycle.
VI.6 - Storage
D
Keep the MD in a sterilisation sachet in a clean and dry
place.
In the event of breakdown, please contact your approved
distributor or our after-sales department directly.
Repairs must only be carried out by an approved repairer
E
or by Anthogyr After Sales Department, only with Anthogyr
original replacement parts.
For all revisions or repairs, the MD must be returned complete
and sterile with proof of sterility. It must be accompanied by
a document outlining the problem and showing the complete
contact details of the user.
The replacement of removable parts is covered for 7 years.
14
W
which are single use) must be sterilised
VII.REPAIR