Once the entire time has passed, the measurement is shown as a
curve on the display.
Note: If the PURGE START: AUTO setting has been activated in
the Sipper options menu, the rinse agent is pumped in automatically
after measurement.
Maintenance
Cleaning
C A U T I O N
Personal injury hazard. Only qualified experts may perform the tasks described in
this section of the manual.
N O T I C E
There may be residual fluid in the tubes. Always wear suitable protective clothing
and goggles when changing the tubes or when performing maintenance work, in
order to prevent contamination of the skin or eyes.
SIP 10
Clean the SIP 10 with a damp cloth. Do not use any type of solvent (for
example, acetone).
Tray for storage (for DR 6000 only)
If you no longer require the SIP 10 for current measurements, you can
store all of the accessories in the tray:
Cell compartment cover and USB connection cable to the right of
the SIP 10
Fitted tubes and rubber tube guide behind the SIP 10
Pour-Thru Cell in the SIP 10
Pour-Thru Cell
Purge the Pour-Thru Cell before and after every series of tests with
deionized water.
If the Pour-Thru Cell is heavily contaminated, repeat the purge cycle
several times.
Before the Pour-Thru Cell is put into storage, perform the purge cycle
several times with air to remove any residual fluid. Hold the Pour-Thru
Cell with the OUT connection slanted downwards during this procedure.