Guidance and manufacturer's declaration – electromagnetic immunity
The model BASE800 is intended for use in the electromagnetic environment specified below. The
customer or the user of the model BASE800 should assure that it is used in such an environment.
Immunity test
± 6 kV contact
Electrostatic
discharge (ESD)
IEC 61000-4-2
± 8 kV air
± 2 kV for power
Electrical fast
supply lines
transient/burst
± 1 kV for
IEC 61000-4-4
input/output
lines
± 1 kV line(s) to
line(s)
Surge
IEC 61000-4-5
± 2 kV line(s) to
earth
<5 %
(>95% dip in UT)
for 0,5 cycle
interruptions and
40 % UT
voltage
(60 % dip in UT)
variations
for 5 cycles
on power supply
input lines
70 % UT
(30 % dip in UT)
IEC 61000-4-11
for 25 cycles
<5 % UT
(>95% dip in UT)
for 5 sec
Power frequency
(50/60 Hz)
magnetic field
3 A/m
IEC 61000-4-8
NOTE UT is the a.c. mains voltage prior to application of the test level.
IEC 60601
Compliance level
test level
± 6 kV contact
± 8 kV air
± 2 kV for power
supply lines
± 1 kV for input/output
lines
± 1 kV line(s) to line(s)
± 2 kV line(s) to earth
UT
<5 %
(>95% dip in UT)
for 0,5 cycle
40 % UT
(60 % dip in UT)
for 5 cycles
70 % UT
(30 % dip in UT)
for 25 cycles
<5 % UT
(>95% dip in UT)
for 5 sec
3 A/m
Electromagnetic environment –
guidance
Floors should be wood, concrete or
ceramic tile. If floors are covered
with synthetic material, the relative
humidity should be at least 30 %.
Mains power quality should be that
of a typical commercial or hospital
environment.
Mains power quality should be that
of a typical commercial or hospital
environment.
UT
Mains power quality should be that
of a typical commercial or hospital
environment. If the user of the
model BASE800 requires continued
operation during power mains
interruptions, it is recommended
that the model BASE800 be
powered from an uninterruptible
power supply or a battery.
Power frequency magnetic fields
should be at levels characteristic
of a typical location in a typical
commercial or hospital environment.